Measurement uncertainty

Complete a compliant GUM / Monte-Carlo uncertainty evaluation in minutes — no need to read JCGM 100/101 first. Pick a scenario, follow the guided steps, watch the budget update live, and export a professional report.

ILAC · Testing · Preparation · General / Cross-disciplineUpgrade to unlock your context & exportable report →
🧭 Get started in 3 minutes

An uncertainty evaluation = a report of “what was measured + how unsure we are”: list every error source (component) with its uᵢ; the tool combines them, derives k, and writes the report.

Before you start, prepare:① the measurand and unit (e.g. grip force N) ② repeated readings or a calibration certificate (either works) ③ the measurement model (optional)
①
Generate from data
You have raw data / calibration certificates / environment records — list components with guidance, combine live, export the budget.
②
Optimize existing report
You have a colleague’s / old / external report (or just basic info) — components are auto-identified and recomputed per GUM; if none are detected, a final-form evaluation report is generated from your stated values with a typical contribution structure (suggested values for you to confirm).
1Define measurand→
2List components→
3Combine→
4Generate report
Free report is generic (ignores context below)
ILAC · Testing · Preparation · General / Cross-discipline
On upgrade, your country / body / industry / stage enter the report pipeline, citing the right mandatory clauses with tailored examples.
① Your measurand
② Evaluation method
GUMRecommended
JCGM 100:2008 law of propagation; default for linear / weakly non-linear models needing a classic budget table.
Monte Carlo
JCGM 101:2008 numerical simulation; safer for strong non-linearity / asymmetric distributions / complex bounds.
Not sure which method? Answer two questions

Enter each component’s standard uncertainty uᵢ, sensitivity cᵢ and degrees of freedom νᵢ. The tool combines as u_c = √Σ(cᵢ·uᵢ)², computes effective df ν_eff (Welch–Satterthwaite), then derives k from the t-distribution at ν_eff (or you set k manually).

③ Uncertainty components

Where does uncertainty come from? Common sources: repeatability, resolution, calibration certificate, temperature, bias. Click above to add a preset component with sensible defaults.

NameTypeDistuᵢ ⓘcᵢ ⓘνᵢ ⓘ
60% · max
34%
5%
1%

④ Analysis report

GUM · generic

Measurement Uncertainty Analysis Report

Generic GUM report · GUM (JCGM 100:2008) · 2026-10-22

LabKitPro · Laboratory Quality Toolkit Doc No. LKP-MU-20261022-3BFD34 · V1.0 · 2026-10-22

1 · Executive summary

VersionGeneric (context not applied)
MeasurandLength (mm)
MethodGUM (JCGM 100:2008)
u_c0.025827k2
U = k·u_c0.051655U (≤2 s.f.)0.052

2 · Measurement model

Y = Σ cᵢ·xᵢ (default linear model)

The model relates the measurand Y to input quantities xᵢ; cᵢ = ∂Y/∂xᵢ is the sensitivity coefficient, taken as 1 when dimensionless.

3 · Sources & uncertainty budget

# Source of uncertainty Type Distribution Basis of evaluation uᵢcᵢ|cᵢ|·uᵢνᵢ
1 Repeatability (Type A) A normal Type A · 10 repeats (Bessel formula) · df ν=n−1=9 0.020000 1 0.020000 9
2 Reference standard (Type B) B normal Type B · normal (certificate gives U; ÷ k: 95%→2, 99%→2.58) 0.015000 1 0.015000 100
3 Fixed point / calibration (Type B, uniform) B uniform Type B · rectangular (half-width a, equiprobable) → u=a/√3 0.005800 1 0.005800 50
4 Resolution (Type B, uniform) B uniform Type B · rectangular (half-width a, equiprobable) → u=a/√3 0.002900 1 0.002900 50

“Basis of evaluation” states how each uᵢ is obtained (Type A = statistical, Type B = non-statistical); “|cᵢ|·uᵢ” feeds the combination and contribution analysis.

Reference · Type B distribution divisors (GUM / NABL-141)

  • Rectangular (half-width a, equiprobable): u = a/√3
  • Triangular (half-width a, mid more likely): u = a/√6
  • Digital resolution ±d/2: treat as rectangular, u = (d/2)/√3
  • Certificate gives expanded U: u = U/k (k is its coverage factor; 95%→k=2, 99%→k=2.58)
  • Avoid “double-counting”: an effect already captured by the Type A statistical spread must not be re-entered as a Type B component.

4 · Combined standard uncertainty

u_c = √Σ(cᵢ·uᵢ)² = √((1×0.020000)² + (1×0.015000)² + (1×0.005800)² + (1×0.002900)²) = 0.025827

(rounded to 3 s.f.: u_c ≈ 0.0258)

Combination assumptions

① Input estimates are treated as uncorrelated (rᵢⱼ = 0) and combined by RSS; if any two components are strongly correlated (|r| > 0.5, e.g. one reference standard used for both calibration and check), the covariance form (GUM §5.2) must be used and this report re-checked. ② Non-linearity of the model is negligible for small input uncertainties (PTB/EURAMET practice); otherwise validate by MCM (JCGM 101).

5 · Effective degrees of freedom

ν_eff = u_c⁴ / Σ[(cᵢ·uᵢ)⁴ / νᵢ] = 24

Welch–Satterthwaite formula; an unknown component df is treated by its type default (Type A uses actual n−1, Type B treated as very large → approx. infinite).

6 · Coverage factor

k = 2 · Derived from the t-distribution at ν_eff (two-sided 95% confidence); a manually set k overrides this.

7 · Expanded uncertainty

U = k·u_c = 2 × 0.025827 = 0.051655

Rounded to at most two significant figures: U ≈ 0.052 mm.

8 · Report statement

The measured result of the measurand Length is reported as Y = …; the expanded uncertainty is U = 0.052 mm (k = 2, approximately 95% confidence). The result and its uncertainty are evaluated and expressed per GUM.

9 · Component contribution

ComponentShare|cᵢ|·uᵢ
Repeatability (Type A)60.0%0.020000
Reference standard (Type B)33.7%0.015000
Fixed point / calibration (Type B, uniform)5.0%0.005800
Resolution (Type B, uniform)1.3%0.002900

The following components were evaluated and found negligible for this u_c (variance share < 5%): Resolution (Type B, uniform) (1.3%). They are retained in the budget for traceability per GUM 4.1.2, not removed from the model.

10 · Dominant component & recommendations

The expanded uncertainty is dominated by Repeatability (Type A), contributing about 60%.

  • Cutting this component by about half (e.g. more repeats to reduce a Type A component, or a higher-grade reference standard to reduce a Type B component) could bring U from 0.0517 down to about 0.0383 mm.
  • Check each “Basis of evaluation” above: Type B components should cite a valid certificate/manual and use the correct distribution divisor (rectangular √3, triangular √6, normal per certificate); Type A degrees of freedom ν=n−1 must match the number of repeats.
  • Prioritise the top one or two contributors — usually the cheapest way to materially shrink U.

11 · Conformity & reporting requirements (ISO/IEC 17025:2017 · 7.6)

  • Uncertainty is evaluated per GUM (JCGM 100:2008); in calibration, CMC states measured quantity / method / range / U in the scope.
  • Expanded uncertainty is given at approximately 95% confidence with the coverage factor k stated; relative expressions such as ppm/ppb are avoided.
  • Results are reported as “y ± U” with a statement of how the uncertainty was obtained.
  • U is rounded to at most two significant figures; u_c keeps one extra figure.
  • CMC intervals are closed (no open intervals such as 0<U<x); reported uncertainty must not be smaller than the accredited CMC.
  • Testing: uncertainty is evaluated and reported whenever it affects validity of results, client requirements, or conformity decisions.

Linked documents: FR-13 Uncertainty Evaluation Report · FR-38 Uncertainty Evaluation List.

Evaluation & Approval

RoleName / SignatureDate
Evaluated by  
Reviewed by  
Approved by  

Disclaimer: this tool is a preparation aid, not a metrological/calibration certificate. Verify against your national accreditation body’s guidance and the latest standard editions, and take responsibility for the model and its validation.

This result is an internal technical calculation draft, not a calibration or measurement certificate.

⚠ This is a BASIC version generated from the generic ISO/IEC 17025:2017 framework. The paid FULL version — tailored to your country / accreditation body (CNAS, A2LA, UKAS, NABL, …) / industry and your uploaded documents — is substantially deeper and field-ready. Download a sample below, or unlock the full deliverable.
BASIC (current free preview)
• Generic ISO/IEC 17025 skeleton
• Generic sample — none of your personalization inputs
• Not tailored to country / body / industry
• Watermarked on-screen preview
FULL (paid deliverable)
• Generic skeleton + your body’s mandatory clauses / prohibited terms
• Your country / body / industry, etc. woven in — expanded per your personalization inputs
• Tailored to your selected context
• Editable, watermark-free Word (.docx)
⚠ Watermarked preview, view-only, not exportable. Unlock for a context-linked report you can export to Word.
  • Editable Word (.docx)
  • Watermark-free
  • Aligned to your shared context
Download sample
❓ Beginner FAQ

What is uncertainty, really?
It is “how unsure the measurement is”. If you measure 100 N with U = 2 N, the true value is within 98–102 N with about 95% confidence.

Type A vs Type B?
Type A = computed from repeated readings (data speaks); Type B = taken from certificates/handbooks/experience (e.g. U÷k on a calibration certificate). Both feed the same combination formula — neither is “better”.

How do I read U in the report?
Results are written “y ± U”; k = 2 means about 95% confidence. U keeps at most 2 significant figures — rounding is handled automatically per GUM.

Free vs paid?
Free: full calculation + online preview (watermarked). Paid: your accreditation body / industry / stage context is injected, watermark-free Word export, and clause-level conformity notes.

This tool performs the GUM combination algebraically and an approximate Monte Carlo simulation in the browser, and re-computes existing reports against GUM rules. It is a preparation aid, not a metrological certificate. Verify against your national body’s guidance and the latest standard editions.